*** MIDAS logfile (03SEPpl1.2) *** Mo, 06 Dec 2004 18:22:05 Page 1 *** MIDAS logfile (vers. 03SEPpl1.2) *** Mo, 06 Dec 2004 18:22:21 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#45 TAKEN ON 041206 IN normcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 349.99 stddev = 2.06 FIERA temperature = 273.15 K There were 0 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_normcd_45_041206 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 3.08 1660.7 1660.9 855.1 0.10 0.08 1.94 2 6.16 3315.6 3315.8 1673.4 -0.10 -0.10 1.98 3 12.31 6630.3 6631.1 3354.1 -0.11 -0.11 1.98 4 18.47 9948.1 9948.5 5008.6 -0.09 -0.09 1.99 5 27.70 14924.9 14926.4 7416.1 -0.07 -0.06 2.01 6 36.93 19911.5 19912.2 9822.4 -0.01 -0.01 2.03 7 46.16 24895.4 24894.9 12091.6 0.01 0.01 2.06 8 61.54 33223.5 33222.8 16008.1 0.10 0.10 2.08 9 76.93 41507.9 41511.8 19671.6 0.05 0.06 2.11 10 92.31 49837.4 49840.0 23069.5 0.11 0.11 2.16 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 3574 to 107257 e- Average count rate measured: 539 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 2.15 +/- 0.02 e-/ADU RON = 4.43 +/- 0.05 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = -0.004 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.11/-0.11 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay