*** MIDAS logfile (03SEPpl1.2) *** Su, 28 Nov 2004 17:16:01 Page 1 *** MIDAS logfile (vers. 03SEPpl1.2) *** Su, 28 Nov 2004 17:16:17 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#45 TAKEN ON 041128 IN normcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 350.23 stddev = 2.07 FIERA temperature = 273.15 K There were 0 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_normcd_45_041128 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 3.08 1646.8 1646.5 847.0 0.11 0.11 1.94 2 6.16 3283.9 3283.6 1656.0 -0.15 -0.14 1.98 3 12.31 6569.9 6570.1 3321.4 -0.09 -0.09 1.98 4 18.47 9853.7 9854.0 4971.9 -0.10 -0.10 1.98 5 27.70 14786.0 14785.7 7259.1 -0.06 -0.06 2.04 6 36.93 19722.8 19723.2 9512.9 -0.01 -0.01 2.07 7 46.16 24662.2 24662.2 11981.6 0.02 0.02 2.06 8 61.54 32906.1 32906.5 15938.8 0.10 0.09 2.06 9 76.93 41131.5 41131.3 19566.0 0.09 0.09 2.10 10 92.31 49355.7 49354.3 22998.1 0.09 0.08 2.15 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 3520 to 105504 e- Average count rate measured: 534 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 2.14 +/- 0.02 e-/ADU RON = 4.43 +/- 0.04 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = -0.002 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.11/-0.15 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay