*** MIDAS logfile (03SEPpl1.2) *** Sa, 13 Nov 2004 23:12:11 Page 1 *** MIDAS logfile (vers. 03SEPpl1.2) *** Sa, 13 Nov 2004 23:12:27 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#45 TAKEN ON 041113 IN normcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 350.61 stddev = 2.07 FIERA temperature = 273.15 K There were 0 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_normcd_45_041113 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 3.08 1682.6 1681.8 881.1 0.09 0.08 1.91 2 6.16 3356.1 3355.9 1728.8 -0.13 -0.12 1.94 3 12.31 6713.3 6713.2 3360.2 -0.10 -0.10 2.00 4 18.47 10069.7 10069.6 5040.1 -0.09 -0.09 2.00 5 27.70 15109.4 15109.9 7537.2 -0.05 -0.05 2.00 6 36.93 20154.5 20154.5 9820.4 -0.01 -0.01 2.05 7 46.16 25202.3 25203.4 12258.8 0.03 0.03 2.06 8 61.54 33624.1 33626.6 16024.3 0.09 0.10 2.10 9 76.93 42021.8 42026.2 19549.9 0.07 0.08 2.15 10 92.31 50434.9 50437.4 23232.7 0.09 0.09 2.17 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 3662 to 109800 e- Average count rate measured: 546 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 2.18 +/- 0.02 e-/ADU RON = 4.51 +/- 0.05 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = -0.002 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.10/-0.13 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay