*** MIDAS logfile (03SEPpl1.2) *** Fr, 05 Nov 2004 13:15:51 Page 1 *** MIDAS logfile (vers. 03SEPpl1.2) *** Fr, 05 Nov 2004 13:16:07 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#45 TAKEN ON 041105 IN normcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 350.87 stddev = 2.08 FIERA temperature = 273.15 K There were 0 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_normcd_45_041105 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 3.08 1634.9 1634.3 844.3 0.11 0.12 1.94 2 6.16 3259.4 3259.2 1845.2 -0.16 -0.15 1.77 3 12.31 6521.8 6521.9 3256.2 -0.09 -0.09 2.00 4 18.47 9781.5 9781.0 4901.6 -0.10 -0.10 2.00 5 27.70 14678.1 14677.3 7328.4 -0.05 -0.06 2.00 6 36.93 19577.7 19579.0 9461.1 -0.02 -0.01 2.07 7 46.16 24482.8 24483.3 12100.3 0.03 0.03 2.02 8 61.54 32661.7 32662.7 15821.1 0.09 0.09 2.06 9 76.93 40835.7 40833.1 19186.6 0.11 0.10 2.13 10 92.31 48992.5 48991.2 22878.8 0.09 0.08 2.14 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 3513 to 105280 e- Average count rate measured: 530 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 2.15 +/- 0.02 e-/ADU RON = 4.47 +/- 0.04 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = -0.002 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.12/-0.15 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay