*** MIDAS logfile (03SEPpl1.2) *** Mo, 01 Nov 2004 15:16:27 Page 1 *** MIDAS logfile (vers. 03SEPpl1.2) *** Mo, 01 Nov 2004 15:16:42 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#45 TAKEN ON 041101 IN normcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 350.66 stddev = 2.07 FIERA temperature = 273.15 K There were 0 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_normcd_45_041101 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 3.08 1700.2 1700.2 900.8 0.12 0.10 1.89 2 6.16 3389.4 3389.1 1718.9 -0.16 -0.17 1.97 3 12.31 6783.3 6782.8 3408.2 -0.08 -0.08 1.99 4 18.47 10171.3 10172.0 5088.3 -0.10 -0.09 2.00 5 27.70 15263.6 15264.0 7545.2 -0.05 -0.04 2.02 6 36.93 20359.8 20358.0 10002.5 -0.01 -0.01 2.04 7 46.16 25461.5 25461.2 12250.7 0.04 0.04 2.08 8 61.54 33961.1 33960.8 16260.1 0.08 0.08 2.09 9 76.93 42466.1 42462.0 20076.4 0.11 0.11 2.12 10 92.31 50931.3 50930.8 23538.2 0.06 0.06 2.16 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 3670 to 109943 e- Average count rate measured: 551 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 2.16 +/- 0.02 e-/ADU RON = 4.47 +/- 0.04 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = -0.001 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.11/-0.16 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay