*** MIDAS logfile (03SEPpl1.2) *** Fr, 08 Oct 2004 14:08:44 Page 1 *** MIDAS logfile (vers. 03SEPpl1.2) *** Fr, 08 Oct 2004 14:09:01 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#45 TAKEN ON 041008 IN normcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 350.78 stddev = 2.07 FIERA temperature = 273.15 K There were 0 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_normcd_45_041008 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 3.08 1657.2 1656.9 857.3 0.11 0.11 1.93 2 6.16 3303.9 3304.1 1678.8 -0.15 -0.15 1.97 3 12.31 6611.6 6611.8 3387.4 -0.08 -0.08 1.95 4 18.47 9914.5 9914.4 4959.9 -0.10 -0.10 2.00 5 27.70 14877.6 14878.2 7381.7 -0.06 -0.05 2.02 6 36.93 19845.7 19845.3 9556.9 -0.01 -0.01 2.08 7 46.16 24817.2 24817.3 11974.0 0.03 0.03 2.07 8 61.54 33105.9 33107.0 15814.5 0.08 0.09 2.09 9 76.93 41393.7 41391.4 19716.4 0.11 0.10 2.10 10 92.31 49652.9 49653.7 23334.6 0.07 0.07 2.13 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 3535 to 105934 e- Average count rate measured: 538 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 2.13 +/- 0.01 e-/ADU RON = 4.42 +/- 0.02 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = -0.001 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.11/-0.15 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay