*** MIDAS logfile (03SEPpl1.2) *** Fr, 24 Sep 2004 20:09:05 Page 1 *** MIDAS logfile (vers. 03SEPpl1.2) *** Fr, 24 Sep 2004 20:09:21 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#45 TAKEN ON 040924 IN normcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 350.33 stddev = 2.08 FIERA temperature = 273.15 K There were 0 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_normcd_45_040924 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 3.08 1661.0 1660.7 852.9 0.10 0.09 1.95 2 6.16 3314.1 3314.4 1681.5 -0.12 -0.11 1.97 3 12.31 6628.9 6628.3 3332.8 -0.09 -0.10 1.99 4 18.47 9942.8 9942.7 4951.5 -0.10 -0.10 2.01 5 27.70 14918.2 14918.4 7331.4 -0.06 -0.06 2.03 6 36.93 19901.7 19902.2 9840.4 -0.01 -0.01 2.02 7 46.16 24884.5 24884.2 12201.9 0.02 0.02 2.04 8 61.54 33201.8 33205.0 15848.3 0.09 0.10 2.09 9 76.93 41493.9 41492.1 19781.0 0.07 0.07 2.10 10 92.31 49803.4 49807.5 23053.9 0.09 0.10 2.16 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 3567 to 106976 e- Average count rate measured: 539 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 2.15 +/- 0.02 e-/ADU RON = 4.46 +/- 0.05 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = -0.002 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.10/-0.11 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay