*** MIDAS logfile (03SEPpl1.2) *** Fr, 10 Sep 2004 13:52:35 Page 1 *** MIDAS logfile (vers. 03SEPpl1.2) *** Fr, 10 Sep 2004 13:52:51 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#45 TAKEN ON 040910 IN normcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 350.51 stddev = 2.08 FIERA temperature = 273.15 K There were 0 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_normcd_45_040910 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 3.08 1725.6 1725.7 899.1 0.11 0.11 1.92 2 6.16 3442.3 3442.4 1749.6 -0.14 -0.14 1.97 3 12.31 6887.0 6887.7 3496.8 -0.09 -0.09 1.97 4 18.47 10329.7 10329.6 5153.6 -0.10 -0.10 2.00 5 27.70 15499.9 15500.4 7729.9 -0.06 -0.06 2.01 6 36.93 20676.1 20676.9 10014.0 -0.01 -0.01 2.06 7 46.16 25857.9 25856.5 12692.4 0.04 0.03 2.04 8 61.54 34494.7 34494.4 16479.2 0.09 0.09 2.09 9 76.93 43119.2 43117.7 20246.1 0.09 0.09 2.13 10 92.31 51733.3 51732.7 23730.6 0.07 0.07 2.18 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 3750 to 112426 e- Average count rate measured: 560 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 2.17 +/- 0.03 e-/ADU RON = 4.52 +/- 0.05 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = -0.002 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.11/-0.14 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay