*** MIDAS logfile (03SEPpl1.2) *** Fr, 03 Sep 2004 18:21:25 Page 1 *** MIDAS logfile (vers. 03SEPpl1.2) *** Fr, 03 Sep 2004 18:21:41 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#45 TAKEN ON 040903 IN normcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 350.34 stddev = 2.07 FIERA temperature = 273.15 K There were 0 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_normcd_45_040903 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 3.08 1722.9 1723.3 889.9 0.08 0.10 1.94 2 6.16 3437.5 3437.3 1745.9 -0.12 -0.13 1.97 3 12.31 6875.4 6875.2 3443.4 -0.10 -0.10 2.00 4 18.47 10313.1 10312.2 5113.1 -0.09 -0.10 2.02 5 27.70 15474.4 15474.0 7730.1 -0.05 -0.06 2.00 6 36.93 20641.9 20642.8 10050.8 -0.00 -0.00 2.05 7 46.16 25811.1 25812.3 12425.8 0.03 0.03 2.08 8 61.54 34436.0 34439.4 16547.2 0.09 0.10 2.08 9 76.93 43037.1 43037.0 20312.5 0.08 0.07 2.12 10 92.31 51648.7 51647.2 23812.1 0.08 0.08 2.17 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 3723 to 111586 e- Average count rate measured: 559 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 2.16 +/- 0.02 e-/ADU RON = 4.47 +/- 0.04 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = -0.001 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.10/-0.13 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay