*** MIDAS logfile (03SEPpl1.2) *** Tu, 24 Aug 2004 13:38:39 Page 1 *** MIDAS logfile (vers. 03SEPpl1.2) *** Tu, 24 Aug 2004 13:38:56 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#45 TAKEN ON 040824 IN normcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 349.75 stddev = 2.05 FIERA temperature = 273.15 K There were 0 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_normcd_45_040824 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 3.08 1722.6 1722.5 902.5 0.10 0.10 1.91 2 6.16 3434.9 3434.7 1750.3 -0.15 -0.15 1.96 3 12.31 6871.8 6872.0 3454.7 -0.09 -0.09 1.99 4 18.47 10305.8 10305.9 5225.6 -0.10 -0.09 1.97 5 27.70 15466.3 15465.0 7669.0 -0.04 -0.05 2.02 6 36.93 20628.2 20628.0 10104.2 -0.00 -0.01 2.04 7 46.16 25794.5 25795.9 12499.3 0.03 0.04 2.06 8 61.54 34411.1 34412.2 16396.2 0.09 0.09 2.10 9 76.93 43013.0 43013.5 20264.3 0.09 0.09 2.12 10 92.31 51601.7 51601.7 23999.5 0.06 0.06 2.15 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 3714 to 111255 e- Average count rate measured: 559 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 2.16 +/- 0.02 e-/ADU RON = 4.42 +/- 0.04 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = -0.000 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.10/-0.15 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay