*** MIDAS logfile (03SEPpl1.2) *** Su, 08 Aug 2004 13:35:18 Page 1 *** MIDAS logfile (vers. 03SEPpl1.2) *** Su, 08 Aug 2004 13:35:34 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#45 TAKEN ON 040808 IN normcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 349.96 stddev = 2.05 FIERA temperature = 273.15 K There were 0 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_normcd_45_040808 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 3.08 1634.3 1634.2 846.6 0.11 0.11 1.93 2 6.16 3259.0 3258.7 1665.9 -0.14 -0.15 1.96 3 12.31 6519.7 6520.5 3281.5 -0.09 -0.08 1.99 4 18.47 9778.8 9777.7 4910.0 -0.09 -0.10 1.99 5 27.70 14673.1 14671.6 7240.2 -0.05 -0.06 2.03 6 36.93 19571.7 19572.1 9673.6 -0.01 -0.01 2.02 7 46.16 24473.9 24474.5 11987.1 0.03 0.03 2.04 8 61.54 32649.3 32648.7 15744.5 0.08 0.08 2.07 9 76.93 40817.7 40814.8 19114.8 0.10 0.09 2.14 10 92.31 48973.6 48977.5 22826.2 0.08 0.09 2.15 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 3515 to 105347 e- Average count rate measured: 530 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 2.15 +/- 0.02 e-/ADU RON = 4.41 +/- 0.04 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = -0.001 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.11/-0.14 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay