*** MIDAS logfile (vers. 02SEPpl1.3) *** Fr, 30 Jul 2004 16:43:56 Page 1 *** MIDAS logfile (vers. 02SEPpl1.3) *** Fr, 30 Jul 2004 16:44:14 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#45 TAKEN ON 040730 IN normcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 349.78 stddev = 2.09 FIERA temperature = 273.15 K There were 0 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_normcd_45_040730 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 3.08 1665.5 1664.6 900.2 0.15 0.13 1.85 2 6.16 3316.6 3316.8 1749.3 -0.22 -0.20 1.90 3 12.31 6641.8 6642.0 3393.1 -0.07 -0.05 1.96 4 18.47 9956.5 9955.3 5074.6 -0.12 -0.12 1.96 5 27.70 14944.7 14943.7 7459.0 -0.05 -0.04 2.00 6 36.93 19931.1 19930.4 9768.4 -0.02 -0.02 2.04 7 46.16 24932.6 24930.5 12266.8 0.06 0.06 2.03 8 61.54 33243.3 33241.9 15904.0 0.06 0.06 2.09 9 76.93 41598.6 41587.3 19717.9 0.17 0.15 2.11 10 92.31 49852.6 49849.6 23474.4 0.04 0.04 2.12 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 3559 to 106570 e- Average count rate measured: 540 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 2.14 +/- 0.02 e-/ADU RON = 4.47 +/- 0.03 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = -0.001 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.16/-0.21 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay