*** MIDAS logfile (vers. 02SEPpl1.3) *** Mo, 26 Jul 2004 20:18:43 Page 1 *** MIDAS logfile (vers. 02SEPpl1.3) *** Mo, 26 Jul 2004 20:19:01 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#45 TAKEN ON 040726 IN normcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 350.10 stddev = 2.07 FIERA temperature = 273.15 K There were 0 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_normcd_45_040726 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 3.08 1704.5 1704.0 904.0 0.11 0.09 1.89 2 6.16 3397.5 3398.2 1748.0 -0.17 -0.15 1.94 3 12.31 6797.9 6798.6 3381.6 -0.10 -0.09 2.01 4 18.47 10195.4 10195.9 5155.5 -0.10 -0.10 1.98 5 27.70 15300.3 15299.7 7614.9 -0.05 -0.05 2.01 6 36.93 20409.3 20408.6 10098.0 -0.00 -0.01 2.02 7 46.16 25520.1 25520.2 12287.6 0.03 0.03 2.08 8 61.54 34047.6 34048.1 16268.4 0.10 0.10 2.09 9 76.93 42561.0 42557.7 19919.2 0.10 0.09 2.14 10 92.31 51062.4 51062.8 23751.4 0.08 0.08 2.15 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 3682 to 110328 e- Average count rate measured: 553 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 2.16 +/- 0.02 e-/ADU RON = 4.47 +/- 0.04 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = -0.000 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.10/-0.16 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay