*** MIDAS logfile (vers. 02SEPpl1.3) *** Tu, 20 Jul 2004 13:28:11 Page 1 *** MIDAS logfile (vers. 02SEPpl1.3) *** Tu, 20 Jul 2004 13:28:29 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#45 TAKEN ON 040720 IN normcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 349.75 stddev = 2.07 FIERA temperature = 273.15 K There were 0 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_normcd_45_040720 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 3.08 1661.8 1662.1 865.8 0.11 0.11 1.92 2 6.16 3312.9 3312.3 1692.9 -0.17 -0.18 1.96 3 12.31 6628.8 6629.7 3343.3 -0.10 -0.08 1.98 4 18.47 9942.0 9941.7 4939.0 -0.10 -0.10 2.01 5 27.70 14918.4 14918.1 7399.9 -0.05 -0.05 2.02 6 36.93 19899.4 19899.4 9693.2 -0.01 -0.01 2.05 7 46.16 24887.2 24884.7 12108.6 0.04 0.04 2.06 8 61.54 33195.0 33193.8 16098.5 0.08 0.08 2.06 9 76.93 41510.1 41504.3 19794.3 0.12 0.11 2.10 10 92.31 49784.5 49783.4 23288.2 0.07 0.07 2.14 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 3541 to 106066 e- Average count rate measured: 539 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 2.13 +/- 0.02 e-/ADU RON = 4.41 +/- 0.04 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = -0.000 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.12/-0.17 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay