*** MIDAS logfile (vers. 02SEPpl1.3) *** Su, 11 Jul 2004 17:45:19 Page 1 *** MIDAS logfile (vers. 02SEPpl1.3) *** Su, 11 Jul 2004 17:45:37 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#45 TAKEN ON 040711 IN normcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 349.61 stddev = 2.07 FIERA temperature = 273.15 K There were 0 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_normcd_45_040711 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 3.08 1734.5 1734.7 899.9 0.09 0.09 1.93 2 6.16 3459.8 3460.2 1778.1 -0.14 -0.14 1.95 3 12.31 6921.6 6922.2 3492.4 -0.09 -0.09 1.98 4 18.47 10380.5 10381.3 5263.0 -0.10 -0.10 1.97 5 27.70 15578.5 15579.1 7570.9 -0.04 -0.04 2.06 6 36.93 20780.4 20779.9 10139.6 -0.00 -0.00 2.05 7 46.16 25984.9 25985.2 12749.4 0.04 0.03 2.04 8 61.54 34664.7 34664.7 16684.0 0.09 0.09 2.08 9 76.93 43331.5 43330.1 20387.7 0.09 0.09 2.13 10 92.31 51984.0 51984.0 23977.2 0.07 0.06 2.17 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 3749 to 112358 e- Average count rate measured: 563 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 2.16 +/- 0.02 e-/ADU RON = 4.47 +/- 0.05 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = -0.001 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.09/-0.14 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay