*** MIDAS logfile (vers. 02SEPpl1.3) *** Sa, 03 Jul 2004 17:32:42 Page 1 *** MIDAS logfile (vers. 02SEPpl1.3) *** Sa, 03 Jul 2004 17:33:00 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#45 TAKEN ON 040703 IN normcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 349.66 stddev = 2.04 FIERA temperature = 273.15 K There were 0 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_normcd_45_040703 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 3.08 1661.5 1660.9 864.1 0.09 0.10 1.92 2 6.16 3314.0 3313.3 1670.4 -0.14 -0.14 1.98 3 12.31 6629.7 6629.7 3318.5 -0.09 -0.09 2.00 4 18.47 9943.6 9942.9 5028.3 -0.10 -0.10 1.98 5 27.70 14920.3 14920.8 7403.2 -0.05 -0.05 2.02 6 36.93 19901.8 19903.1 9733.3 -0.01 -0.01 2.04 7 46.16 24889.0 24889.1 12202.1 0.03 0.03 2.04 8 61.54 33203.3 33202.3 15873.1 0.09 0.09 2.09 9 76.93 41504.5 41506.3 19671.1 0.09 0.09 2.11 10 92.31 49800.4 49800.3 23049.8 0.08 0.08 2.16 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 3578 to 107256 e- Average count rate measured: 539 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 2.15 +/- 0.02 e-/ADU RON = 4.40 +/- 0.04 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = -0.002 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.10/-0.14 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay