*** MIDAS logfile (vers. 02SEPpl1.3) *** Th, 17 Jun 2004 13:38:28 Page 1 *** MIDAS logfile (vers. 02SEPpl1.3) *** Th, 17 Jun 2004 13:38:47 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#45 TAKEN ON 040617 IN normcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 349.51 stddev = 2.06 FIERA temperature = 273.15 K There were 0 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_normcd_45_040617 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 3.08 1738.1 1738.2 908.9 0.09 0.11 1.91 2 6.16 3467.3 3467.1 1763.5 -0.13 -0.12 1.97 3 12.31 6934.0 6935.5 3467.4 -0.11 -0.10 2.00 4 18.47 10403.7 10402.7 5280.5 -0.08 -0.10 1.97 5 27.70 15609.8 15609.1 7827.3 -0.05 -0.06 1.99 6 36.93 20822.1 20823.8 10182.9 -0.00 -0.00 2.04 7 46.16 26036.3 26036.4 12579.2 0.03 0.02 2.07 8 61.54 34739.7 34740.2 16637.4 0.10 0.10 2.09 9 76.93 43409.8 43410.4 20463.0 0.07 0.07 2.12 10 92.31 52097.9 52099.3 24420.7 0.08 0.08 2.13 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 3727 to 111717 e- Average count rate measured: 564 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 2.14 +/- 0.02 e-/ADU RON = 4.41 +/- 0.03 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = -0.002 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.11/-0.13 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay