*** MIDAS logfile (vers. 02SEPpl1.3) *** Mo, 31 May 2004 21:09:20 Page 1 *** MIDAS logfile (vers. 02SEPpl1.3) *** Mo, 31 May 2004 21:09:37 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#45 TAKEN ON 040531 IN normcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 349.26 stddev = 2.05 FIERA temperature = 273.15 K There were 0 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_normcd_45_040531 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 3.08 1690.1 1690.2 898.6 0.11 0.12 1.88 2 6.16 3370.0 3369.8 1738.8 -0.15 -0.16 1.94 3 12.31 6744.1 6744.4 3367.3 -0.08 -0.07 2.00 4 18.47 10112.9 10112.7 5051.8 -0.10 -0.10 2.00 5 27.70 15176.4 15177.0 7530.2 -0.05 -0.04 2.02 6 36.93 20243.5 20242.7 9885.2 -0.01 -0.01 2.05 7 46.16 25317.1 25314.3 12406.5 0.04 0.03 2.04 8 61.54 33767.9 33767.6 16027.4 0.08 0.08 2.11 9 76.93 42214.3 42215.8 19828.6 0.09 0.10 2.13 10 92.31 50639.0 50638.3 23612.8 0.06 0.06 2.14 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 3641 to 109097 e- Average count rate measured: 548 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 2.15 +/- 0.02 e-/ADU RON = 4.42 +/- 0.04 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = -0.001 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.12/-0.16 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay