*** MIDAS logfile (vers. 02SEPpl1.3) *** Su, 23 May 2004 22:53:21 Page 1 *** MIDAS logfile (vers. 02SEPpl1.3) *** Su, 23 May 2004 22:53:38 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#45 TAKEN ON 040523 IN normcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 349.77 stddev = 2.06 FIERA temperature = 273.15 K There were 0 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_normcd_45_040523 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 3.08 1778.4 1778.2 916.0 0.10 0.10 1.94 2 6.16 3547.2 3546.6 1797.0 -0.13 -0.15 1.97 3 12.31 7096.2 7096.2 3508.9 -0.09 -0.09 2.02 4 18.47 10642.8 10643.3 5279.5 -0.10 -0.09 2.02 5 27.70 15968.7 15969.8 7918.1 -0.06 -0.05 2.02 6 36.93 21302.7 21303.0 10388.7 -0.01 -0.00 2.05 7 46.16 26639.2 26638.6 13087.2 0.03 0.03 2.04 8 61.54 35537.5 35539.4 17140.9 0.09 0.10 2.07 9 76.93 44424.0 44419.5 21040.0 0.09 0.09 2.11 10 92.31 53294.2 53295.6 25009.4 0.07 0.07 2.13 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 3788 to 113531 e- Average count rate measured: 577 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 2.13 +/- 0.02 e-/ADU RON = 4.40 +/- 0.04 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = -0.001 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.10/-0.14 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay