*** MIDAS logfile (vers. 02SEPpl1.3) *** Mo, 17 May 2004 17:15:54 Page 1 *** MIDAS logfile (vers. 02SEPpl1.3) *** Mo, 17 May 2004 17:16:11 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#45 TAKEN ON 040517 IN normcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 349.68 stddev = 2.08 FIERA temperature = 273.15 K There were 0 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_normcd_45_040517 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 3.08 1682.9 1683.2 870.6 0.14 0.13 1.93 2 6.16 3353.3 3353.2 1734.0 -0.20 -0.20 1.93 3 12.31 6714.7 6714.5 3324.6 -0.06 -0.06 2.02 4 18.47 10065.6 10065.1 5010.2 -0.12 -0.11 2.01 5 27.69 15107.9 15106.6 7431.3 -0.04 -0.05 2.03 6 36.93 20149.4 20149.3 9807.1 -0.02 -0.01 2.05 7 46.16 25204.3 25201.6 12244.4 0.05 0.05 2.06 8 61.54 33609.1 33607.1 16198.6 0.07 0.07 2.07 9 76.93 42043.8 42038.6 19791.3 0.14 0.14 2.12 10 92.31 50396.2 50393.1 23735.3 0.03 0.04 2.12 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 3589 to 107473 e- Average count rate measured: 546 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 2.13 +/- 0.01 e-/ADU RON = 4.44 +/- 0.03 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = -0.001 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.14/-0.20 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay