*** MIDAS logfile (vers. 02SEPpl1.3) *** Su, 16 May 2004 14:41:10 Page 1 *** MIDAS logfile (vers. 02SEPpl1.3) *** Su, 16 May 2004 14:41:27 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#45 TAKEN ON 040516 IN normcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 349.81 stddev = 2.08 FIERA temperature = 273.15 K There were 0 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_normcd_45_040516 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 3.08 1695.3 1695.2 907.5 0.16 0.15 1.87 2 6.16 3373.5 3373.6 1726.3 -0.26 -0.24 1.95 3 12.31 6758.9 6758.8 3362.3 -0.04 -0.03 2.01 4 18.47 10127.7 10126.2 5082.8 -0.13 -0.13 1.99 5 27.70 15204.5 15202.8 7446.9 -0.03 -0.03 2.04 6 36.93 20277.1 20275.1 9934.8 -0.01 -0.01 2.04 7 46.16 25367.5 25363.8 12498.7 0.08 0.08 2.03 8 61.54 33809.7 33805.4 16186.8 0.04 0.04 2.09 9 76.93 42334.2 42322.8 19968.4 0.21 0.20 2.12 10 92.31 50685.7 50678.1 23708.6 -0.01 -0.02 2.14 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 3632 to 108569 e- Average count rate measured: 549 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 2.14 +/- 0.02 e-/ADU RON = 4.45 +/- 0.04 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = 0.002 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.21/-0.25 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay