*** MIDAS logfile (vers. 02SEPpl1.3) *** Fr, 07 May 2004 17:38:41 Page 1 *** MIDAS logfile (vers. 02SEPpl1.3) *** Fr, 07 May 2004 17:38:59 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#45 TAKEN ON 040507 IN normcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 349.60 stddev = 2.05 FIERA temperature = 273.15 K There were 0 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_normcd_45_040507 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 3.08 1714.6 1714.3 913.6 0.13 0.12 1.88 2 6.16 3414.9 3415.3 1751.2 -0.22 -0.21 1.95 3 12.31 6839.6 6839.4 3466.2 -0.05 -0.05 1.97 4 18.47 10251.6 10250.2 5091.9 -0.12 -0.13 2.01 5 27.70 15388.0 15386.0 7500.9 -0.04 -0.05 2.05 6 36.93 20524.2 20521.1 10006.9 -0.01 -0.01 2.05 7 46.16 25674.6 25671.5 12484.9 0.07 0.07 2.06 8 61.54 34228.6 34228.1 16374.2 0.06 0.07 2.09 9 76.93 42828.6 42824.8 20482.9 0.16 0.16 2.09 10 92.31 51323.8 51318.8 23634.0 0.02 0.02 2.17 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 3691 to 110499 e- Average count rate measured: 556 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 2.15 +/- 0.02 e-/ADU RON = 4.42 +/- 0.05 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = -0.000 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.16/-0.22 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay