*** MIDAS logfile (vers. 02SEPpl1.3) *** Fr, 30 Apr 2004 15:53:11 Page 1 *** MIDAS logfile (vers. 02SEPpl1.3) *** Fr, 30 Apr 2004 15:53:28 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#45 TAKEN ON 040430 IN normcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 349.57 stddev = 2.06 FIERA temperature = 273.15 K There were 0 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_normcd_45_040430 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 3.08 1694.7 1694.6 891.3 0.19 0.18 1.90 2 6.16 3368.3 3368.0 1742.0 -0.34 -0.34 1.93 3 12.31 6757.6 6756.0 3376.2 0.02 0.02 2.00 4 18.47 10116.8 10115.3 5084.1 -0.16 -0.15 1.99 5 27.70 15192.4 15190.0 7551.7 -0.03 -0.02 2.01 6 36.93 20258.3 20253.4 9952.2 -0.02 -0.01 2.04 7 46.16 25355.8 25348.1 12142.0 0.12 0.12 2.09 8 61.54 33768.4 33760.2 16248.4 0.01 0.01 2.08 9 76.93 42330.4 42311.3 19953.8 0.29 0.28 2.12 10 92.31 50604.4 50593.3 23327.4 -0.09 -0.08 2.17 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 3667 to 109479 e- Average count rate measured: 549 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 2.16 +/- 0.02 e-/ADU RON = 4.45 +/- 0.05 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = 0.003 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.28/-0.34 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay