*** MIDAS logfile (vers. 02SEPpl1.3) *** Su, 25 Apr 2004 16:25:51 Page 1 *** MIDAS logfile (vers. 02SEPpl1.3) *** Su, 25 Apr 2004 16:26:07 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#45 TAKEN ON 040425 IN normcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 349.47 stddev = 2.05 FIERA temperature = 273.15 K There were 0 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_normcd_45_040425 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 3.08 1778.1 1778.5 915.1 0.09 0.10 1.94 2 6.16 3548.0 3548.0 1814.8 -0.12 -0.13 1.96 3 12.31 7097.3 7096.9 3540.0 -0.10 -0.10 2.00 4 18.47 10645.4 10646.0 5368.2 -0.10 -0.10 1.98 5 27.70 15974.6 15975.8 7880.5 -0.05 -0.05 2.03 6 36.93 21310.6 21309.4 10349.6 -0.00 -0.01 2.06 7 46.16 26647.0 26647.0 13050.2 0.03 0.03 2.04 8 61.54 35553.8 35554.7 17013.6 0.10 0.10 2.09 9 76.93 44428.8 44429.9 20870.7 0.07 0.07 2.13 10 92.31 53320.2 53322.1 24453.8 0.08 0.08 2.18 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 3861 to 115780 e- Average count rate measured: 577 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 2.17 +/- 0.03 e-/ADU RON = 4.46 +/- 0.05 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = -0.002 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.10/-0.13 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay