*** MIDAS logfile (vers. 02SEPpl1.3) *** Tu, 20 Apr 2004 17:11:30 Page 1 *** MIDAS logfile (vers. 02SEPpl1.3) *** Tu, 20 Apr 2004 17:11:47 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#45 TAKEN ON 040420 IN normcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 349.35 stddev = 2.05 FIERA temperature = 273.15 K There were 0 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_normcd_45_040420 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 3.08 1713.7 1714.0 893.6 0.11 0.11 1.92 2 6.16 3416.2 3416.5 1747.2 -0.18 -0.17 1.96 3 12.31 6837.2 6838.1 3440.0 -0.09 -0.08 1.99 4 18.47 10253.6 10254.8 5061.8 -0.11 -0.10 2.03 5 27.70 15389.1 15388.9 7699.3 -0.05 -0.05 2.00 6 36.93 20528.1 20527.9 9985.1 0.00 -0.00 2.06 7 46.16 25672.1 25670.6 12376.9 0.05 0.04 2.07 8 61.54 34242.4 34241.2 16436.6 0.09 0.08 2.08 9 76.93 42814.8 42813.7 20356.0 0.12 0.11 2.10 10 92.31 51351.9 51350.2 23646.4 0.07 0.06 2.17 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 3697 to 110780 e- Average count rate measured: 556 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 2.16 +/- 0.02 e-/ADU RON = 4.43 +/- 0.05 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = -0.001 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.12/-0.18 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay