*** MIDAS logfile (vers. 02SEPpl1.3) *** Mo, 12 Apr 2004 16:43:17 Page 1 *** MIDAS logfile (vers. 02SEPpl1.3) *** Mo, 12 Apr 2004 16:43:34 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#45 TAKEN ON 040412 IN normcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 349.28 stddev = 2.05 FIERA temperature = 273.15 K There were 0 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_normcd_45_040412 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 3.08 1721.4 1721.3 898.9 0.09 0.08 1.91 2 6.16 3434.9 3435.1 1767.9 -0.11 -0.11 1.94 3 12.31 6870.2 6869.9 3425.7 -0.10 -0.11 2.01 4 18.47 10306.4 10306.4 5226.4 -0.09 -0.09 1.97 5 27.70 15463.5 15463.8 7701.4 -0.05 -0.06 2.01 6 36.93 20628.7 20628.6 10049.6 -0.00 -0.01 2.05 7 46.16 25791.1 25791.9 12436.1 0.02 0.02 2.07 8 61.54 34417.5 34421.4 16279.6 0.10 0.11 2.11 9 76.93 42998.7 43001.8 20162.5 0.05 0.05 2.13 10 92.31 51624.5 51625.8 23912.0 0.10 0.10 2.16 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 3732 to 111936 e- Average count rate measured: 559 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 2.17 +/- 0.02 e-/ADU RON = 4.45 +/- 0.04 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = -0.002 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.11/-0.11 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay