*** MIDAS logfile (vers. 02SEPpl1.3) *** We, 07 Apr 2004 19:03:22 Page 1 *** MIDAS logfile (vers. 02SEPpl1.3) *** We, 07 Apr 2004 19:03:39 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#45 TAKEN ON 040407 IN normcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 349.33 stddev = 2.06 FIERA temperature = 273.15 K There were 0 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_normcd_45_040407 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 3.08 1720.2 1720.1 905.9 0.11 0.09 1.90 2 6.16 3430.2 3431.1 1760.0 -0.14 -0.12 1.95 3 12.31 6862.3 6862.1 3449.4 -0.10 -0.10 1.99 4 18.47 10293.1 10293.8 5200.0 -0.10 -0.09 1.98 5 27.70 15445.2 15445.1 7640.1 -0.05 -0.06 2.02 6 36.93 20603.9 20601.0 10081.6 -0.00 -0.02 2.04 7 46.16 25760.5 25762.2 12336.2 0.02 0.03 2.09 8 61.54 34374.3 34375.0 16544.3 0.10 0.10 2.08 9 76.93 42955.1 42958.9 20134.2 0.07 0.08 2.13 10 92.31 51556.1 51558.8 23730.9 0.09 0.09 2.17 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 3734 to 111909 e- Average count rate measured: 558 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 2.17 +/- 0.02 e-/ADU RON = 4.48 +/- 0.05 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = -0.001 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.10/-0.13 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay