*** MIDAS logfile (vers. 02SEPpl1.3) *** We, 31 Mar 2004 17:55:52 Page 1 *** MIDAS logfile (vers. 02SEPpl1.3) *** We, 31 Mar 2004 17:56:10 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#45 TAKEN ON 040331 IN normcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 349.48 stddev = 2.04 FIERA temperature = 273.15 K There were 0 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_normcd_45_040331 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 3.08 1781.3 1781.2 913.2 0.10 0.09 1.95 2 6.16 3552.1 3553.8 1828.8 -0.16 -0.12 1.94 3 12.31 7108.2 7107.2 3573.3 -0.09 -0.10 1.99 4 18.47 10661.3 10662.0 5390.9 -0.09 -0.09 1.98 5 27.70 15997.2 15996.8 7901.8 -0.05 -0.05 2.02 6 36.93 21339.7 21341.3 10416.9 0.00 0.00 2.05 7 46.16 26682.9 26683.0 12967.3 0.03 0.03 2.06 8 61.54 35599.4 35601.6 17063.0 0.10 0.10 2.09 9 76.93 44491.3 44489.7 20919.2 0.08 0.07 2.13 10 92.31 53384.7 53388.6 24606.6 0.07 0.07 2.17 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 3858 to 115624 e- Average count rate measured: 578 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 2.17 +/- 0.02 e-/ADU RON = 4.42 +/- 0.04 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = -0.001 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.10/-0.14 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay