*** MIDAS logfile (vers. 02SEPpl1.3) *** Tu, 09 Mar 2004 13:32:36 Page 1 *** MIDAS logfile (vers. 02SEPpl1.3) *** Tu, 09 Mar 2004 13:32:53 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#45 TAKEN ON 040309 IN normcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 349.11 stddev = 2.05 FIERA temperature = 273.15 K There were 0 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_normcd_45_040309 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 3.08 1748.9 1748.3 914.1 0.12 0.10 1.91 2 6.16 3487.1 3486.6 1810.1 -0.15 -0.15 1.93 3 12.31 6976.0 6976.5 3489.9 -0.10 -0.09 2.00 4 18.47 10463.8 10463.6 5245.8 -0.09 -0.09 1.99 5 27.70 15702.3 15701.1 7703.5 -0.05 -0.05 2.04 6 36.93 20944.2 20943.8 10200.1 -0.01 -0.01 2.05 7 46.16 26190.8 26192.4 12577.9 0.03 0.04 2.08 8 61.54 34937.6 34940.2 16805.9 0.08 0.09 2.08 9 76.93 43675.4 43672.1 20345.1 0.09 0.09 2.15 10 92.31 52395.9 52395.7 24345.4 0.06 0.07 2.15 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 3780 to 113263 e- Average count rate measured: 567 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 2.16 +/- 0.02 e-/ADU RON = 4.43 +/- 0.04 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = -0.001 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.11/-0.15 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay