*** MIDAS logfile (vers. 02SEPpl1.3) *** Mo, 01 Mar 2004 17:44:32 Page 1 *** MIDAS logfile (vers. 02SEPpl1.3) *** Mo, 01 Mar 2004 17:44:49 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#45 TAKEN ON 040301 IN normcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 349.35 stddev = 2.04 FIERA temperature = 273.15 K There were 0 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_normcd_45_040301 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 3.08 1739.3 1739.1 903.1 0.11 0.10 1.93 2 6.16 3469.4 3469.9 1764.0 -0.12 -0.11 1.97 3 12.31 6939.3 6939.8 3499.4 -0.11 -0.10 1.98 4 18.47 10410.3 10410.3 5230.5 -0.09 -0.09 1.99 5 27.70 15619.2 15621.0 7795.1 -0.06 -0.05 2.00 6 36.93 20835.3 20836.7 10378.4 -0.01 -0.01 2.01 7 46.16 26052.2 26053.7 12710.0 0.02 0.02 2.05 8 61.54 34764.3 34765.9 16737.9 0.10 0.10 2.08 9 76.93 43438.1 43438.9 20602.0 0.06 0.06 2.11 10 92.31 52141.6 52143.2 24275.0 0.09 0.09 2.15 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 3730 to 111830 e- Average count rate measured: 564 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 2.14 +/- 0.02 e-/ADU RON = 4.37 +/- 0.04 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = -0.002 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.11/-0.12 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay