*** MIDAS logfile (vers. 02SEPpl1.3) *** Su, 08 Feb 2004 14:22:58 Page 1 *** MIDAS logfile (vers. 02SEPpl1.3) *** Su, 08 Feb 2004 14:23:52 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#45 TAKEN ON 040208 IN normcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 349.01 stddev = 2.03 FIERA temperature = 273.15 K There were 0 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_normcd_45_040208 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 3.08 1764.5 1765.0 908.5 0.10 0.09 1.94 2 6.16 3520.4 3520.7 1803.4 -0.13 -0.13 1.95 3 12.31 7043.1 7041.9 3571.5 -0.09 -0.10 1.97 4 18.47 10563.6 10563.7 5246.2 -0.09 -0.09 2.01 5 27.70 15851.0 15850.1 7725.4 -0.05 -0.05 2.05 6 36.93 21144.3 21144.8 10326.0 -0.01 -0.00 2.05 7 46.16 26437.7 26437.6 12727.7 0.02 0.03 2.08 8 61.54 35275.6 35276.3 16882.1 0.10 0.10 2.09 9 76.93 44079.7 44086.4 20719.9 0.06 0.08 2.13 10 92.31 52904.0 52906.4 24343.3 0.08 0.09 2.17 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 3825 to 114673 e- Average count rate measured: 573 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 2.17 +/- 0.02 e-/ADU RON = 4.40 +/- 0.04 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = -0.002 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.10/-0.13 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay