*** MIDAS logfile (vers. 02SEPpl1.3) *** Su, 01 Feb 2004 18:51:21 Page 1 *** MIDAS logfile (vers. 02SEPpl1.3) *** Su, 01 Feb 2004 18:52:23 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#45 TAKEN ON 040201 IN normcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 348.95 stddev = 2.03 FIERA temperature = 273.15 K There were 0 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_normcd_45_040201 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 3.08 1833.4 1833.8 965.9 0.09 0.08 1.90 2 6.16 3660.6 3660.9 1841.8 -0.08 -0.08 1.99 3 12.31 7318.6 7319.0 3715.0 -0.12 -0.12 1.97 4 18.47 10982.7 10983.5 5469.3 -0.08 -0.08 2.01 5 27.70 16475.7 16477.3 8086.5 -0.06 -0.06 2.04 6 36.93 21981.3 21984.2 10657.5 -0.00 0.00 2.06 7 46.16 27479.2 27480.0 13350.9 0.01 0.00 2.06 8 61.54 36681.8 36684.6 17393.8 0.13 0.13 2.11 9 76.93 45805.1 45810.3 21095.2 0.02 0.03 2.17 10 92.31 55008.4 55014.2 25536.4 0.10 0.10 2.15 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 3988 to 119656 e- Average count rate measured: 595 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 2.18 +/- 0.02 e-/ADU RON = 4.42 +/- 0.04 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = -0.003 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.13/-0.12 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay