*** MIDAS logfile (vers. 02SEPpl1.3) *** Fr, 16 Jan 2004 13:30:20 Page 1 *** MIDAS logfile (vers. 02SEPpl1.3) *** Fr, 16 Jan 2004 13:31:09 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#45 TAKEN ON 040116 IN normcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 349.28 stddev = 2.05 FIERA temperature = 273.15 K There were 0 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_normcd_45_040116 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 3.08 1762.5 1762.8 933.5 0.12 0.13 1.89 2 6.16 3514.9 3514.2 1784.5 -0.15 -0.16 1.97 3 12.31 7033.1 7032.6 3529.6 -0.09 -0.10 1.99 4 18.47 10548.9 10548.4 5188.0 -0.10 -0.10 2.03 5 27.69 15829.0 15829.8 7859.8 -0.05 -0.05 2.01 6 36.93 21115.6 21115.6 10403.4 -0.00 -0.00 2.03 7 46.16 26403.6 26404.5 12704.6 0.03 0.04 2.08 8 61.54 35227.9 35224.8 16854.4 0.10 0.09 2.09 9 76.93 44030.9 44031.0 20643.9 0.09 0.09 2.13 10 92.31 52821.1 52824.1 23939.4 0.06 0.07 2.21 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 3859 to 115633 e- Average count rate measured: 572 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 2.19 +/- 0.03 e-/ADU RON = 4.49 +/- 0.06 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = -0.002 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.12/-0.15 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay