*** MIDAS logfile (vers. 02SEPpl1.3) *** Fr, 09 Jan 2004 16:37:56 Page 1 *** MIDAS logfile (vers. 02SEPpl1.3) *** Fr, 09 Jan 2004 16:38:47 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#45 TAKEN ON 040109 IN normcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 349.36 stddev = 2.05 FIERA temperature = 273.15 K There were 0 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_normcd_45_040109 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 3.08 1822.2 1822.2 943.8 0.10 0.10 1.93 2 6.16 3634.4 3634.8 1840.8 -0.13 -0.12 1.97 3 12.31 7270.2 7270.7 3658.4 -0.10 -0.09 1.99 4 18.47 10906.2 10905.4 5449.5 -0.08 -0.09 2.00 5 27.70 16362.6 16363.6 8163.2 -0.05 -0.05 2.00 6 36.93 21829.3 21829.4 10700.4 0.01 0.00 2.04 7 46.16 27294.6 27295.2 13257.1 0.04 0.03 2.06 8 61.54 36415.3 36415.9 17395.7 0.10 0.10 2.09 9 76.93 45497.4 45502.7 21543.1 0.05 0.06 2.11 10 92.31 54607.1 54608.0 25129.5 0.07 0.07 2.17 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 3941 to 118109 e- Average count rate measured: 591 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 2.16 +/- 0.02 e-/ADU RON = 4.44 +/- 0.05 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = -0.001 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.10/-0.13 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay