*** MIDAS logfile (vers. 02SEPpl1.3) *** Fr, 02 Jan 2004 16:29:03 Page 1 *** MIDAS logfile (vers. 02SEPpl1.3) *** Fr, 02 Jan 2004 16:29:54 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#45 TAKEN ON 040102 IN normcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 349.41 stddev = 2.05 FIERA temperature = 273.15 K There were 0 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_normcd_45_040102 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 3.08 1751.6 1751.2 898.3 0.10 0.10 1.95 2 6.16 3493.0 3492.8 1794.8 -0.14 -0.15 1.95 3 12.31 6987.0 6986.9 3498.9 -0.10 -0.11 2.00 4 18.47 10481.1 10481.0 5239.3 -0.09 -0.09 2.00 5 27.70 15724.8 15726.2 7739.3 -0.06 -0.05 2.03 6 36.93 20977.9 20978.1 10200.7 -0.01 -0.01 2.06 7 46.16 26231.1 26229.5 12822.8 0.03 0.02 2.05 8 61.54 34998.2 35002.2 16775.9 0.10 0.11 2.09 9 76.93 43738.4 43736.7 20749.1 0.08 0.07 2.11 10 92.31 52495.1 52499.0 24383.8 0.09 0.10 2.15 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 3760 to 112691 e- Average count rate measured: 568 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 2.15 +/- 0.02 e-/ADU RON = 4.40 +/- 0.04 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = -0.001 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.10/-0.14 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay