*** MIDAS logfile (05SEPpl1.2) *** Sa, 09 Feb 2008 13:33:00 Page 1 *** MIDAS logfile (vers. 05SEPpl1.2) *** Sa, 09 Feb 2008 13:33:14 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#46 TAKEN ON 080209 IN fastcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 283.64 stddev = 1.99 FIERA temperature = 273.15 K There were 0 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_fastcd_46_080209 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 0.55 1604.3 1605.5 781.5 0.12 0.03 2.05 2 1.09 3190.0 3189.5 1628.2 -0.20 -0.26 1.96 3 2.17 6396.8 6399.8 3186.4 0.07 0.08 2.01 4 3.25 9580.6 9579.9 4799.5 -0.06 -0.07 2.00 5 4.88 14382.0 14384.8 7391.8 0.04 0.06 1.95 6 6.50 19159.1 19164.3 9572.9 -0.06 -0.03 2.00 7 8.13 23968.5 23972.6 11831.1 0.03 0.06 2.03 8 10.84 31947.7 31948.1 16118.9 0.01 0.02 1.98 9 13.55 39940.2 39938.7 19604.5 0.02 0.04 2.04 10 16.25 47932.6 47934.9 24086.6 0.03 0.06 1.99 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 3221 to 96209 e- Average count rate measured: 2948 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 2.01 +/- 0.01 e-/ADU RON = 3.99 +/- 0.03 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = -0.002 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.10/-0.23 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay