*** MIDAS logfile (05SEPpl1.2) *** We, 30 Jan 2008 15:34:49 Page 1 *** MIDAS logfile (vers. 05SEPpl1.2) *** We, 30 Jan 2008 15:35:03 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#46 TAKEN ON 080130 IN fastcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 283.69 stddev = 1.97 FIERA temperature = 273.15 K There were 0 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_fastcd_46_080130 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 0.55 1601.5 1602.3 788.8 0.08 0.13 2.03 2 1.09 3185.3 3184.9 1514.8 -0.20 -0.20 2.10 3 2.17 6390.2 6387.9 3153.5 0.10 0.07 2.03 4 3.25 9563.9 9564.1 4811.0 -0.08 -0.08 1.99 5 4.88 14358.6 14359.8 7311.4 0.02 0.04 1.96 6 6.50 19132.9 19133.4 9600.6 -0.05 -0.05 1.99 7 8.13 23938.1 23937.3 11817.3 0.05 0.05 2.03 8 10.84 31896.4 31894.1 15941.6 -0.00 -0.01 2.00 9 13.55 39887.4 39886.6 20140.1 0.04 0.04 1.98 10 16.25 47861.5 47855.5 24661.0 0.04 0.02 1.94 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 3135 to 93674 e- Average count rate measured: 2944 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 1.96 +/- 0.02 e-/ADU RON = 3.86 +/- 0.03 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = -0.002 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.11/-0.20 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay