*** MIDAS logfile (05SEPpl1.2) *** Th, 20 Dec 2007 16:12:00 Page 1 *** MIDAS logfile (vers. 05SEPpl1.2) *** Th, 20 Dec 2007 16:12:14 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#46 TAKEN ON 071220 IN fastcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 284.84 stddev = 2.08 FIERA temperature = 273.15 K There were 0 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_fastcd_46_071220 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 0.55 1603.9 1604.0 815.5 0.12 0.12 1.97 2 1.09 3187.8 3190.9 1555.6 -0.23 -0.17 2.05 3 2.17 6398.2 6395.6 3178.3 0.10 0.08 2.01 4 3.25 9573.7 9574.9 4802.7 -0.09 -0.09 1.99 5 4.88 14377.4 14379.7 7236.6 0.04 0.04 1.99 6 6.50 19154.3 19157.0 9770.5 -0.06 -0.05 1.96 7 8.13 23967.7 23967.7 12198.4 0.05 0.05 1.96 8 10.84 31934.3 31934.0 15872.6 -0.01 -0.02 2.01 9 13.55 39943.0 39937.7 19956.0 0.05 0.03 2.00 10 16.25 47916.8 47914.0 24607.9 0.03 0.01 1.95 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 3158 to 94353 e- Average count rate measured: 2948 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 1.97 +/- 0.02 e-/ADU RON = 4.10 +/- 0.03 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = -0.002 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.12/-0.20 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay