*** MIDAS logfile (05SEPpl1.2) *** Fr, 19 Oct 2007 13:17:57 Page 1 *** MIDAS logfile (vers. 05SEPpl1.2) *** Fr, 19 Oct 2007 13:18:11 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#46 TAKEN ON 071019 IN fastcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 285.00 stddev = 1.98 FIERA temperature = 273.15 K There were 0 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_fastcd_46_071019 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 0.55 1604.3 1604.7 817.6 0.01 0.04 1.96 2 1.09 3193.7 3193.8 1585.3 -0.18 -0.19 2.01 3 2.17 6401.7 6401.5 3183.8 0.10 0.10 2.01 4 3.25 9581.4 9576.2 4867.6 -0.07 -0.11 1.97 5 4.88 14387.3 14382.7 7152.5 0.06 0.04 2.01 6 6.50 19164.6 19167.1 9916.3 -0.04 -0.03 1.93 7 8.13 23974.6 23977.2 12090.0 0.05 0.06 1.98 8 10.84 31946.2 31948.3 16505.5 -0.00 0.01 1.94 9 13.55 39949.7 39949.8 19452.4 0.04 0.04 2.05 10 16.25 47932.0 47933.3 23728.6 0.03 0.04 2.02 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 3240 to 96797 e- Average count rate measured: 2948 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 2.02 +/- 0.02 e-/ADU RON = 4.01 +/- 0.05 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = -0.001 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.10/-0.19 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay