*** MIDAS logfile (05SEPpl1.2) *** We, 19 Sep 2007 16:35:21 Page 1 *** MIDAS logfile (vers. 05SEPpl1.2) *** We, 19 Sep 2007 16:35:35 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#46 TAKEN ON 070919 IN fastcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 285.15 stddev = 1.97 FIERA temperature = 273.15 K There were 0 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_fastcd_46_070919 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 0.55 1601.0 1604.0 793.1 0.04 0.12 2.02 2 1.09 3192.2 3189.2 1598.2 -0.13 -0.19 2.00 3 2.17 6396.7 6396.6 3179.6 0.09 0.08 2.01 4 3.25 9576.1 9576.9 4713.8 -0.07 -0.06 2.03 5 4.88 14373.6 14376.2 7266.4 0.02 0.03 1.98 6 6.50 19153.7 19149.8 9652.9 -0.06 -0.07 1.98 7 8.13 23961.0 23960.2 11651.7 0.03 0.03 2.06 8 10.84 31934.1 31936.4 15617.9 0.00 0.00 2.04 9 13.55 39925.5 39927.8 19859.2 0.02 0.02 2.01 10 16.25 47919.6 47923.7 23581.3 0.05 0.05 2.03 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 3256 to 97358 e- Average count rate measured: 2947 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 2.03 +/- 0.01 e-/ADU RON = 4.00 +/- 0.02 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = -0.002 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.10/-0.16 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay