*** MIDAS logfile (05SEPpl1.2) *** We, 29 Aug 2007 16:41:53 Page 1 *** MIDAS logfile (vers. 05SEPpl1.2) *** We, 29 Aug 2007 16:42:07 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#46 TAKEN ON 070829 IN fastcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 285.10 stddev = 1.96 FIERA temperature = 273.15 K There were 0 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_fastcd_46_070829 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 0.55 1611.5 1612.2 797.3 0.12 0.04 2.02 2 1.09 3208.0 3207.0 1627.4 -0.16 -0.25 1.97 3 2.17 6428.4 6431.9 3289.9 0.08 0.09 1.95 4 3.25 9624.6 9624.9 4756.3 -0.08 -0.08 2.02 5 4.88 14451.1 14448.8 7063.2 0.03 0.03 2.05 6 6.50 19251.4 19252.7 9826.0 -0.07 -0.04 1.96 7 8.13 24086.4 24087.6 12114.0 0.03 0.06 1.99 8 10.84 32095.2 32095.9 15874.6 -0.02 0.01 2.02 9 13.55 40138.8 40142.8 20125.7 0.03 0.07 1.99 10 16.25 48161.6 48161.7 24280.8 0.03 0.06 1.98 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 3210 to 95904 e- Average count rate measured: 2962 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 1.99 +/- 0.01 e-/ADU RON = 3.90 +/- 0.02 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = -0.002 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.11/-0.20 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay