*** MIDAS logfile (05SEPpl1.2) *** Th, 21 Jun 2007 16:28:33 Page 1 *** MIDAS logfile (vers. 05SEPpl1.2) *** Th, 21 Jun 2007 16:28:47 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#46 TAKEN ON 070621 IN fastcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 285.07 stddev = 1.98 FIERA temperature = 273.15 K There were 0 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_fastcd_46_070621 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 0.55 1640.6 1640.3 810.5 0.08 0.07 2.02 2 1.09 3268.2 3266.0 1680.1 -0.16 -0.18 1.95 3 2.17 6549.9 6552.8 3265.7 0.09 0.11 2.01 4 3.25 9808.4 9807.8 4878.9 -0.05 -0.06 2.01 5 4.88 14722.9 14719.3 7205.6 0.03 0.02 2.04 6 6.50 19614.1 19612.8 9791.4 -0.06 -0.06 2.00 7 8.13 24537.1 24539.2 12051.5 0.03 0.04 2.04 8 10.84 32707.2 32703.1 16596.7 0.01 0.00 1.97 9 13.55 40886.8 40885.3 20057.2 0.01 0.01 2.04 10 16.25 49068.7 49068.2 23790.0 0.03 0.03 2.06 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 3356 to 100370 e- Average count rate measured: 3018 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 2.05 +/- 0.02 e-/ADU RON = 4.06 +/- 0.04 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = -0.002 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.10/-0.17 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay