*** MIDAS logfile (05SEPpl1.2) *** Su, 13 May 2007 16:21:37 Page 1 *** MIDAS logfile (vers. 05SEPpl1.2) *** Su, 13 May 2007 16:21:51 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#46 TAKEN ON 070513 IN fastcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 284.97 stddev = 1.96 FIERA temperature = 273.15 K There were 0 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_fastcd_46_070513 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 0.55 1657.6 1656.6 847.8 0.13 0.05 1.96 2 1.09 3298.2 3298.4 1656.3 -0.17 -0.17 1.99 3 2.17 6613.2 6613.6 3359.5 0.10 0.12 1.97 4 3.25 9897.2 9894.3 4998.5 -0.08 -0.09 1.98 5 4.88 14857.8 14862.0 7550.6 0.02 0.05 1.97 6 6.50 19795.3 19797.5 9756.1 -0.07 -0.05 2.03 7 8.13 24773.2 24771.1 12432.0 0.05 0.05 1.99 8 10.84 33007.2 33005.8 16848.7 -0.01 -0.01 1.96 9 13.55 41271.3 41273.6 20644.3 0.02 0.03 2.00 10 16.25 49518.2 49521.6 24131.9 0.01 0.02 2.05 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 3358 to 100348 e- Average count rate measured: 3046 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 2.03 +/- 0.02 e-/ADU RON = 3.98 +/- 0.04 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = -0.002 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.12/-0.17 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay