*** MIDAS logfile (05SEPpl1.2) *** Sa, 24 Mar 2007 16:13:48 Page 1 *** MIDAS logfile (vers. 05SEPpl1.2) *** Sa, 24 Mar 2007 16:14:02 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#46 TAKEN ON 070324 IN fastcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 284.80 stddev = 1.97 FIERA temperature = 273.15 K There were 0 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_fastcd_46_070324 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 0.55 1693.5 1697.2 837.1 0.05 0.03 2.02 2 1.09 3380.1 3380.1 1653.3 -0.08 -0.14 2.04 3 2.17 6775.6 6774.8 3434.1 0.12 0.10 1.97 4 3.25 10140.0 10138.7 5079.1 -0.08 -0.09 2.00 5 4.88 15228.4 15227.0 7586.1 0.04 0.05 2.01 6 6.50 20283.1 20286.7 10019.0 -0.08 -0.04 2.02 7 8.13 25383.5 25383.1 12496.9 0.04 0.06 2.03 8 10.84 33812.6 33816.8 15945.1 -0.04 -0.01 2.12 9 13.55 42311.3 42295.2 20969.3 0.06 0.05 2.02 10 16.25 50732.5 50729.6 25524.7 -0.02 0.00 1.99 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 3423 to 102426 e- Average count rate measured: 3122 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 2.02 +/- 0.03 e-/ADU RON = 3.98 +/- 0.05 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = -0.003 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.11/-0.11 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay