*** MIDAS logfile (05SEPpl1.2) *** Fr, 16 Feb 2007 15:51:37 Page 1 *** MIDAS logfile (vers. 05SEPpl1.2) *** Fr, 16 Feb 2007 15:51:50 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#46 TAKEN ON 070216 IN fastcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 284.74 stddev = 2.01 FIERA temperature = 273.15 K There were 0 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_fastcd_46_070216 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 0.55 1713.5 1716.4 857.6 0.09 0.18 2.00 2 1.09 3403.0 3406.5 1731.7 -0.31 -0.27 1.97 3 2.17 6846.9 6844.4 3360.6 0.31 0.28 2.04 4 3.25 10219.9 10219.2 5160.3 -0.13 -0.15 1.98 5 4.88 15341.8 15341.7 7790.2 -0.03 -0.04 1.97 6 6.50 20442.0 20442.8 10372.6 -0.10 -0.11 1.97 7 8.13 25570.8 25568.6 12894.1 -0.02 -0.04 1.98 8 10.84 34081.6 34079.6 17161.7 -0.04 -0.06 1.99 9 13.55 42720.8 42722.8 21184.4 0.24 0.23 2.02 10 16.25 51130.5 51133.1 25226.0 -0.02 -0.02 2.03 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 3464 to 103290 e- Average count rate measured: 3147 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 2.02 +/- 0.01 e-/ADU RON = 4.06 +/- 0.02 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = -0.001 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.29/-0.29 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay