*** MIDAS logfile (05SEPpl1.2) *** Su, 04 Feb 2007 15:26:15 Page 1 *** MIDAS logfile (vers. 05SEPpl1.2) *** Su, 04 Feb 2007 15:26:29 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#46 TAKEN ON 070204 IN fastcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 284.65 stddev = 1.99 FIERA temperature = 273.15 K There were 0 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_fastcd_46_070204 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 0.55 1721.5 1724.4 852.1 0.02 0.11 2.02 2 1.09 3427.5 3425.4 1718.9 -0.18 -0.21 1.99 3 2.17 6870.3 6867.2 3484.5 0.11 0.07 1.97 4 3.25 10281.1 10280.0 5207.6 -0.07 -0.08 1.97 5 4.88 15435.2 15435.6 7877.9 0.04 0.04 1.96 6 6.50 20563.3 20566.2 10239.0 -0.04 -0.03 2.01 7 8.13 25727.7 25727.6 12735.9 0.06 0.05 2.02 8 10.84 34281.2 34281.3 17357.2 0.01 0.00 1.98 9 13.55 42854.5 42857.5 21134.2 0.02 0.02 2.03 10 16.25 51426.2 51424.9 24685.1 0.02 0.02 2.08 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 3547 to 105863 e- Average count rate measured: 3163 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 2.06 +/- 0.02 e-/ADU RON = 4.09 +/- 0.05 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = -0.001 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.11/-0.19 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay