*** MIDAS logfile (05SEPpl1.2) *** We, 06 Dec 2006 14:43:16 Page 1 *** MIDAS logfile (vers. 05SEPpl1.2) *** We, 06 Dec 2006 14:43:30 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#46 TAKEN ON 061206 IN fastcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 284.00 stddev = 1.97 FIERA temperature = 273.15 K There were 0 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_fastcd_46_061206 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 0.55 1698.8 1699.3 845.2 0.09 0.11 2.01 2 1.09 3381.3 3382.2 1726.5 -0.15 -0.12 1.96 3 2.17 6777.7 6774.2 3359.1 0.09 0.06 2.02 4 3.25 10143.8 10143.5 5095.4 -0.08 -0.09 1.99 5 4.88 15232.0 15232.5 7647.1 0.04 0.04 1.99 6 6.50 20293.0 20294.1 10467.4 -0.05 -0.05 1.94 7 8.13 25388.9 25386.3 12851.8 0.04 0.04 1.98 8 10.84 33830.6 33833.9 16572.5 -0.01 -0.00 2.04 9 13.55 42294.9 42294.3 21766.5 0.01 0.00 1.94 10 16.25 50760.6 50756.3 25342.2 0.03 0.01 2.00 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 3378 to 100919 e- Average count rate measured: 3123 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 1.99 +/- 0.02 e-/ADU RON = 3.91 +/- 0.04 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = -0.002 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.10/-0.14 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay