*** MIDAS logfile (05SEPpl1.2) *** Su, 26 Nov 2006 12:27:31 Page 1 *** MIDAS logfile (vers. 05SEPpl1.2) *** Su, 26 Nov 2006 12:27:44 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#46 TAKEN ON 061126 IN fastcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 284.11 stddev = 1.98 FIERA temperature = 273.15 K There were 0 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_fastcd_46_061126 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 0.55 1707.3 1707.2 832.4 0.07 0.04 2.05 2 1.09 3398.8 3402.3 1684.2 -0.21 -0.16 2.02 3 2.17 6817.9 6820.5 3508.1 0.08 0.10 1.94 4 3.25 10209.1 10207.7 5078.4 -0.07 -0.09 2.01 5 4.88 15324.0 15330.7 7634.4 0.02 0.05 2.01 6 6.50 20424.2 20422.1 10313.5 -0.03 -0.04 1.98 7 8.13 25551.6 25547.1 12851.0 0.06 0.04 1.99 8 10.84 34049.1 34044.0 16890.6 0.01 -0.01 2.02 9 13.55 42569.6 42575.8 20542.1 0.03 0.04 2.07 10 16.25 51082.3 51084.0 25305.7 0.03 0.03 2.02 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 3481 to 104154 e- Average count rate measured: 3142 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 2.04 +/- 0.02 e-/ADU RON = 4.03 +/- 0.03 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = -0.002 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.09/-0.19 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay