*** MIDAS logfile (05SEPpl1.2) *** Fr, 10 Nov 2006 15:35:55 Page 1 *** MIDAS logfile (vers. 05SEPpl1.2) *** Fr, 10 Nov 2006 15:36:08 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#46 TAKEN ON 061110 IN fastcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 284.09 stddev = 1.99 FIERA temperature = 273.15 K There were 0 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_fastcd_46_061110 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 0.55 1713.2 1716.2 843.8 0.09 0.06 2.03 2 1.09 3413.8 3411.5 1677.7 -0.11 -0.20 2.03 3 2.17 6841.6 6839.4 3404.1 0.11 0.08 2.01 4 3.25 10239.7 10239.6 5116.2 -0.07 -0.07 2.00 5 4.88 15373.9 15372.5 7634.7 0.04 0.04 2.01 6 6.50 20478.7 20483.8 10266.6 -0.07 -0.03 1.99 7 8.13 25624.3 25624.9 12931.1 0.04 0.06 1.98 8 10.84 34142.0 34142.7 16648.2 -0.02 0.01 2.05 9 13.55 42687.3 42689.2 21021.3 0.00 0.03 2.03 10 16.25 51219.4 51221.9 25094.3 -0.00 0.03 2.04 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 3499 to 104533 e- Average count rate measured: 3151 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 2.04 +/- 0.01 e-/ADU RON = 4.06 +/- 0.02 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = -0.002 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.09/-0.16 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay