*** MIDAS logfile (05SEPpl1.2) *** Sa, 04 Nov 2006 12:13:29 Page 1 *** MIDAS logfile (vers. 05SEPpl1.2) *** Sa, 04 Nov 2006 12:13:43 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#46 TAKEN ON 061104 IN fastcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 284.26 stddev = 1.99 FIERA temperature = 273.15 K There were 0 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_fastcd_46_061104 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 0.55 1703.7 1705.4 845.8 0.07 0.02 2.01 2 1.09 3387.9 3390.4 1687.2 -0.21 -0.24 2.01 3 2.17 6799.1 6799.7 3299.1 0.11 0.10 2.06 4 3.25 10175.0 10176.3 5080.2 -0.07 -0.07 2.00 5 4.88 15279.4 15279.7 7586.1 0.05 0.06 2.01 6 6.50 20353.0 20351.8 10041.7 -0.06 -0.04 2.03 7 8.13 25466.7 25467.4 13192.6 0.05 0.07 1.93 8 10.84 33932.6 33934.7 16692.8 -0.00 0.02 2.03 9 13.55 42431.4 42431.4 21339.9 0.03 0.05 1.99 10 16.25 50906.3 50905.2 24630.6 0.02 0.03 2.07 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 3466 to 103518 e- Average count rate measured: 3131 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 2.03 +/- 0.03 e-/ADU RON = 4.05 +/- 0.05 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = -0.002 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.11/-0.23 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay